Jfca. Veloso et al., APPLICATION OF A MICROSTRIP GAS COUNTER IN ENERGY-DISPERSIVE X-RAY-FLUORESCENCE ANALYSIS, X-ray spectrometry, 26(4), 1997, pp. 237-243
Performance characteristics of a microstrip gas counter operated as an
x-ray fluorescence spectrometer are reported, Gas amplification as a
function of microstrip anode-cathode voltage was measured, and the bre
akdown threshold voltage was determined in pure xenon. The detector te
mporal stability and the effect of gas purity were assessed, Energy re
solution and linearity, detection efficiency and uniformity of spatial
response in the 2-60 keV x-ray energy range were determined from the
pulse-height distributions of the fluorescence x-ray spectra induced i
n a variety of single- and multi-element sample materials, Energy reso
lution similar to that of conventional proportional counters was achie
ved at 6 keV. (C) 1997 by John Wiley & Sons, Ltd.