On-line monitoring of nitric oxide complexed with porphyrine-bearing biochemical materials by using flow injection with chemiluminescence detection

Citation
Dc. Yao et al., On-line monitoring of nitric oxide complexed with porphyrine-bearing biochemical materials by using flow injection with chemiluminescence detection, ANALYT CHIM, 435(2), 2001, pp. 273-280
Citations number
21
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICA CHIMICA ACTA
ISSN journal
00032670 → ACNP
Volume
435
Issue
2
Year of publication
2001
Pages
273 - 280
Database
ISI
SICI code
0003-2670(20010524)435:2<273:OMONOC>2.0.ZU;2-M
Abstract
An on-line nitric oxide (NO) monitoring method is described employing a flo w injection manifold with chemiluminescence (CL) detection. The method is t ailored for monitoring the NO from its complexes with porphyrin ring-bearin g (PRB) biochemical materials. The method is optimized for sensitivity and detection limit. The method is quite sensitive, simple, rapid, and precise. A sensitivity of 2.63 x 10(5) mV mol(-1) NO, a range of 2.0 x 10(-5) up to 1.0 x 10(-2) mol NO (under the conditions used) a reproducibility of 1.0% over the range 1.0 x 10(-3) up to 1.0 x 10-2 mol NO, a frequency of ca. fiv e samples per minute and a detection limit of 1.0 x 10(-5) mol NO are the a nalytical figures of merit for the proposed method. (C) 2001 Elsevier Scien ce B.V. All rights reserved.