Copper(II)-selective electrode using 2,2 '-dithiodianiline as neutral carrier

Citation
Mb. Gholivand et N. Nozari, Copper(II)-selective electrode using 2,2 '-dithiodianiline as neutral carrier, TALANTA, 54(4), 2001, pp. 597-602
Citations number
36
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
TALANTA
ISSN journal
00399140 → ACNP
Volume
54
Issue
4
Year of publication
2001
Pages
597 - 602
Database
ISI
SICI code
0039-9140(20010530)54:4<597:CEU2'A>2.0.ZU;2-C
Abstract
Poly(vinyl chloride) membrane electrode, that is highly selective and sensi tive to Cu(II) ions, was developed by using 2,2'-dithiodianiline and dibuty l phthalate as carrier and plasticizer, respectively. The electrode exhibit s good potentiometric response for Cu(II) over a wide concentration range ( 5.0 x 10(-2)-7.0 x 10(-7) mol l(-1))with Nernstian slope of 30 +/- 1 mV per decade. The response time of the electrode is 10 s and it has been used fo r a period of one month and exhibits good selectivity towards Cu2+ in compa rison to alkali, alkaline earth, transition and heavy metal ions, with no i nterference caused by Pb2+, Cd2+ and Fe+2 which are known to interfere with many other copper electrodes. (C) 2001 Elsevier Science B.V. All rights re served.