Multicentre collaborative study to calibrate IGF-II by bioassay and immunoassay: establishment of the First WHO Reference Reagent

Citation
B. Rafferty et al., Multicentre collaborative study to calibrate IGF-II by bioassay and immunoassay: establishment of the First WHO Reference Reagent, GROWTH H I, 11(1), 2001, pp. 18-23
Citations number
10
Categorie Soggetti
Endocrinology, Nutrition & Metabolism
Journal title
GROWTH HORMONE & IGF RESEARCH
ISSN journal
10966374 → ACNP
Volume
11
Issue
1
Year of publication
2001
Pages
18 - 23
Database
ISI
SICI code
1096-6374(200102)11:1<18:MCSTCI>2.0.ZU;2-N
Abstract
A preparation of recombinant insulin-like growth factor-II (IGF-II) (NIBSC code 96/538) was compared with local standards in bioassays and immunoassay s by eight laboratories in four countries to assess its suitability for use as a World Health Organisation (WHO) reference reagent. Estimates of relat ive potencies for the bioassays gave a geometric mean of 1.04 (0.94-1.16) m ug of local standard per mug of 96/538. Estimates of relative immunological activities by immunoassay gave a geometric mean of 1.15 (0.94-1.38) mug of local standard per mug of 96/538. The study provided evidence that a commo n standard for rhlGF-II would be helpful and that 96/538 was sufficiently s table to serve as a reference reagent. Accordingly 96/538 was established a s the First WHO Reference Reagent for IGF-II, human, recombinant, and assig ned a unitage of 5000 units per ampoule and on the basis of the immunoassay results a nominal mass content of 5 mug per ampoule. (C) 2001 Harcourt Pub lishers Ltd.