B. Rafferty et al., Multicentre collaborative study to calibrate IGF-II by bioassay and immunoassay: establishment of the First WHO Reference Reagent, GROWTH H I, 11(1), 2001, pp. 18-23
A preparation of recombinant insulin-like growth factor-II (IGF-II) (NIBSC
code 96/538) was compared with local standards in bioassays and immunoassay
s by eight laboratories in four countries to assess its suitability for use
as a World Health Organisation (WHO) reference reagent. Estimates of relat
ive potencies for the bioassays gave a geometric mean of 1.04 (0.94-1.16) m
ug of local standard per mug of 96/538. Estimates of relative immunological
activities by immunoassay gave a geometric mean of 1.15 (0.94-1.38) mug of
local standard per mug of 96/538. The study provided evidence that a commo
n standard for rhlGF-II would be helpful and that 96/538 was sufficiently s
table to serve as a reference reagent. Accordingly 96/538 was established a
s the First WHO Reference Reagent for IGF-II, human, recombinant, and assig
ned a unitage of 5000 units per ampoule and on the basis of the immunoassay
results a nominal mass content of 5 mug per ampoule. (C) 2001 Harcourt Pub
lishers Ltd.