Md. Axelsson et I. Rodushkin, Determination of major and trace elements in sphalerite using laser ablation double focusing sector field ICP-MS, J GEOCHEM E, 72(2), 2001, pp. 81-89
The analytical performance of laser ablation (LA) for the determination of
Co, Fe, Cd, Ag, Mn, Cu and S in sphalerite was evaluated using double focus
ing sector field inductively coupled plasma mass spectrometry (ICP-SFMS). S
amples were collected from Zinkgruvan, situated in the south central Sweden
. The use of Zn for internal standardisation, together with correction for
FeS impurities in sphalerite, allows straightforward quantification without
using external methods for the determination of the actual Zn content. LA-
ICP-SFMS results were compared with data obtained by conventional pneumatic
nebulisation introduction of sample solutions following acid digestion. Go
od agreement between the two methods was obtained for homogeneously distrib
uted elements. For the majority of the elements under consideration, LA-ICP
-SFMS precision was better than 10% RSD. (C) 2001 Elsevier Science B.V. All
rights reserved.