The new measurement system for high-resolution, high-rate gamma -spectromet
ry consists of a 20% n-HPGe detector with a transistor reset preamplifier,
a PLDF-system, and a software implementation of an LFC-multichannel analyze
r allowing the continuous registration of 120 pairs of 16k channel gamma -s
pectra, pile-up and dead-time corrected, and uncorrected [3, 4]. Thus, it i
s possible to obtain the total information contained in the gamma -spectra
of short- and medium-lived radionuclides after a short neutron irradiation.
The new gamma -spectrometry system was tested by activation of a mixture o
f In, Se, and Ba. This enables the optimal analysis of radionuclides by sel
ecting the optimal summation range simultaneously with a quality assurance
of the data by plotting the decay curves of the respective radionuclides.