The effect of graphite tube condition on measured trace Pb concentrations in ETAA studies

Citation
S. Govender et al., The effect of graphite tube condition on measured trace Pb concentrations in ETAA studies, J TR MICROP, 19(2), 2001, pp. 217-223
Citations number
11
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF TRACE AND MICROPROBE TECHNIQUES
ISSN journal
07334680 → ACNP
Volume
19
Issue
2
Year of publication
2001
Pages
217 - 223
Database
ISI
SICI code
0733-4680(2001)19:2<217:TEOGTC>2.0.ZU;2-A
Abstract
This investigation evaluated the variation in the quantitative parameters o f ETAA by comparing data recorded with fired (old) and new graphite tubes ( furnaces) under identical experimental conditions. It was discovered that t he results (for Pb) for a tube that was fired a 100 times varied significan tly with those obtained for a new tube. This led to the inference that it w ould be appropriate to include the replicate number of firings in reported data associated with ETAA investigations.