The sensitivity of GAFChromic dosimetry using a conventional broad band lig
ht source densitometer has significantly been improved twofold using an ine
xpensive red acetate filter overlay during the densitometric measurements.
This thin sheet of red acetate enhances the dosimetric analysis of radiochr
omic blue image distributions recorded on GAFChromic films. The combination
provides higher sensitivity in the optical density measurements than the m
ore expensive He-Ne laser-scanning densitometers. (C) 2001 American Associa
tion of Physicists in Medicine.