The use of an inexpensive red acetate filter to improve the sensitivity ofGAFChromic dosimetry

Citation
Do. Odero et al., The use of an inexpensive red acetate filter to improve the sensitivity ofGAFChromic dosimetry, MED PHYS, 28(7), 2001, pp. 1446-1448
Citations number
10
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
MEDICAL PHYSICS
ISSN journal
00942405 → ACNP
Volume
28
Issue
7
Year of publication
2001
Pages
1446 - 1448
Database
ISI
SICI code
0094-2405(200107)28:7<1446:TUOAIR>2.0.ZU;2-F
Abstract
The sensitivity of GAFChromic dosimetry using a conventional broad band lig ht source densitometer has significantly been improved twofold using an ine xpensive red acetate filter overlay during the densitometric measurements. This thin sheet of red acetate enhances the dosimetric analysis of radiochr omic blue image distributions recorded on GAFChromic films. The combination provides higher sensitivity in the optical density measurements than the m ore expensive He-Ne laser-scanning densitometers. (C) 2001 American Associa tion of Physicists in Medicine.