A new technique for unbiased external ion accumulation in a quadrupole two-dimensional ion trap for electrospray ionization Fourier transform ion cyclotron resonance mass spectrometry
Me. Belov et al., A new technique for unbiased external ion accumulation in a quadrupole two-dimensional ion trap for electrospray ionization Fourier transform ion cyclotron resonance mass spectrometry, RAP C MASS, 15(14), 2001, pp. 1172-1180
External ion accumulation in a two-dimensional (2D) multipole trap has been
shown to increase the sensitivity, dynamic range and duty cycle of a Fouri
er transform ion cyclotron resonance (FTICR) mass spectrometer. However, it
is important that trapped ions be detected without significant bias at lon
ger accumulation times in the external 2D multipole trap. With increasing i
on accumulation time pronounced m/z discrimination was observed when trappi
ng ions in an accumulation quadrupole. In this work we show that superimpos
ing lower rf-amplitude dipolar excitation over the main rf-field in the acc
umulation quadrupole results in disruption of the m/z discrimination and ca
n potentially be used to achieve unbiased external ion accumulation with FT
ICR. Copyright (C) 2001 John Wiley & Sons, Ltd.