M. Garzotti et al., Coupling of a supercritical fluid chromatography system to a hybrid (Q-TOF2) mass spectrometer: on-line accurate mass measurements, RAP C MASS, 15(14), 2001, pp. 1187-1190
This note describes a simple and economical method to couple a supercritica
l fluid chromatography (SFC) system (Berger Instruments, US) with a high-re
solution hybrid mass spectrometer (Q-TOF 2; Micromass, UK). This experiment
al arrangement has three distinct advantages: (1) coupling between the two
systems can be effected without the need for an interface or hardware modif
ications of either system, (ii) this experimental arrangement provides on-l
ine accurate mass SFC/MS measurements which are indispensable for the chara
cterisation of new chemical entities and unknown metabolites, and (iii) the
characteristically fast spectral acquisition rate of the time-of-flight (T
OF) analyser renders the present arrangement an important contribution to f
uture semipreparative fraction collection setups which use mass spectrometr
y as a detector. Copyright (C) 2001 John Wiley & Sons, Ltd.