A novel method for beat-to-beat detection of ventricular late potentials

Citation
Sc. Wu et al., A novel method for beat-to-beat detection of ventricular late potentials, IEEE BIOMED, 48(8), 2001, pp. 931-935
Citations number
13
Categorie Soggetti
Multidisciplinary,"Instrumentation & Measurement
Journal title
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
ISSN journal
00189294 → ACNP
Volume
48
Issue
8
Year of publication
2001
Pages
931 - 935
Database
ISI
SICI code
0018-9294(200108)48:8<931:ANMFBD>2.0.ZU;2-4
Abstract
A novel method for beat-to-beat detection of ventricular late potentials (V LP) from high-resolution electrocardiograms (ECGs) is presented. ECG signal s from the X lead are first filtered using a bandpass filter, and then a ti me-sequence adaptive filter, to improve its signal-to-noise ratio. Eight fe atures are extracted using wavelet transform, from the VLP time-frequency d istribution of the filtered ECG signals, and used as inputs of specially de signed artificial neural network for VLP recognition. The artificial neural network was trained and tested using clinical data, respectively. The resu lts show that the presented method can detect beat-to-beat-based VLP with s ensitivity of 80% and specificity of 77%, and the detection accuracy is 78% .