X-ray fluorescence escape peaks in Ge(Li) detectors

Citation
R. Durak et Y. Ozdemir, X-ray fluorescence escape peaks in Ge(Li) detectors, INSTR SCI T, 29(3), 2001, pp. 185-192
Citations number
12
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
INSTRUMENTATION SCIENCE & TECHNOLOGY
ISSN journal
10739149 → ACNP
Volume
29
Issue
3
Year of publication
2001
Pages
185 - 192
Database
ISI
SICI code
1073-9149(2001)29:3<185:XFEPIG>2.0.ZU;2-9
Abstract
The escape peak-to-photopeak intensity ratios and escape fractions of chara cteristic It X-rays, as a function of incident radiation energy in the rang e 15.75-50 keV and collimator diameter (or area seen by the crystal), are m easured for a Ge(Li) detector. The experimental results are compared with t heoretical predictions.