Ultratrace and isotopic analysis of long-lived radionuclides by double-focusing sector field inductively coupled plasma mass spectrometry using direct liquid sample introduction

Citation
Ja. Mclean et al., Ultratrace and isotopic analysis of long-lived radionuclides by double-focusing sector field inductively coupled plasma mass spectrometry using direct liquid sample introduction, INT J MASS, 208(1-3), 2001, pp. 193-204
Citations number
49
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY
ISSN journal
13873806 → ACNP
Volume
208
Issue
1-3
Year of publication
2001
Pages
193 - 204
Database
ISI
SICI code
1387-3806(20010718)208:1-3<193:UAIAOL>2.0.ZU;2-J
Abstract
This report is concerned with the investigation of double-focusing sector f ield inductively coupled plasma mass spectrometry (DF-ICPMS) for ultratrace and isotopic ratio analysis of long-lived radionuclides (Ra-226, Th-230, T h-232, U-233, Np-237, U-238, and Am-241) using the direct injection high ef ficiency nebulizer (DIHEN). A new shielded torch arrangement, known as the guard electrode, improves relative sensitivity by a factor of six when the DIHEN is used. Absolute sensitivity with the DIHEN is on the order of 1300 (Ra-226) to 1700 (U-238) counts/fg at a solution consumption rate of 5 muL/ min. This is a factor of from three to 20 better than the results obtained by a conventional nebulizer-spray chamber arrangement (e.g., ultrasonic and pneumatic nebulizers). The DIHEN-DF-ICPMS is successfully tested for isoto pe ratio measurements of U-235/U-238 standards and environmental radioactiv e waste solutions. (C) 2001 Elsevier Science B.V.