Videodensitometric HPTLC determination of fleroxacin, sparfloxacin, and cinoxacin in tablets

Citation
D. Kowalczuk et H. Hopkata, Videodensitometric HPTLC determination of fleroxacin, sparfloxacin, and cinoxacin in tablets, J PL CHROM, 14(2), 2001, pp. 126-129
Citations number
11
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JPC-JOURNAL OF PLANAR CHROMATOGRAPHY-MODERN TLC
ISSN journal
09334173 → ACNP
Volume
14
Issue
2
Year of publication
2001
Pages
126 - 129
Database
ISI
SICI code
0933-4173(200103/04)14:2<126:VHDOFS>2.0.ZU;2-A
Abstract
A new, simple, accurate, and rapid high-performance thin-layer chromatograp hic method has been developed for the determination of fleroxacin, sparflox acin, and cinoxacin in tablets. HPTLC was performed on silica gel 60 F-254 plates in horizontal chambers with dichloromethane-isopropyl alcohol-25% NH 3, 4 + 5 + 2, as mobile phase. Detection and quantification were achieved b y means of videodensitometry at lambda = 254 nm. The range of linearity was 0.08-0.48 mug muL(-1) for all the drugs analyzed. The sensitivity of detec tion was 10 ng spot(-1) for cinoxacin and 100 ng spot(-1) for sparfloxacin and fleroxacin. The relative standard deviations were <5.3%. The recovery o f the three antibiotics from tablets ranged from 96 to 105%.