Highly selective and sensitive copper(II) membrane coated graphite electrode based on a recently synthesized Schiff's base

Citation
Mr. Ganjali et al., Highly selective and sensitive copper(II) membrane coated graphite electrode based on a recently synthesized Schiff's base, ANALYT CHIM, 440(2), 2001, pp. 81-87
Citations number
37
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICA CHIMICA ACTA
ISSN journal
00032670 → ACNP
Volume
440
Issue
2
Year of publication
2001
Pages
81 - 87
Database
ISI
SICI code
0003-2670(20010809)440:2<81:HSASCM>2.0.ZU;2-D
Abstract
A PVC membrane electrode based on bis-2-thiophenal propanediamine (TPDA) co ated directly on graphite is described. The electrode exhibits a Nernstian response for Cu2+ over a very wide concentration range (1.0 x 10(-1) to 6.0 x 10(-8) M) with a detection limit of 3.0 x 10(-8) M (2.56 ng ml(-1)). It has a fast response time and can be used for at least 2 months without any major deviation. The proposed sensor revealed very good selectivities for C u2+ over a wide variety of other metal ions and could be used in the pH ran ge of 3.0-7.0. It was successfully used for direct determination of copper in black tea and as an indicator electrode in potentiometric titration of c opper ion. (C) 2001 Elsevier Science B.V All rights reserved.