Rapid on-site determination of chlorobenzene in water samples using ion mobility spectrometry

Citation
H. Borsdorf et al., Rapid on-site determination of chlorobenzene in water samples using ion mobility spectrometry, ANALYT CHIM, 440(1), 2001, pp. 63-70
Citations number
11
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICA CHIMICA ACTA
ISSN journal
00032670 → ACNP
Volume
440
Issue
1
Year of publication
2001
Pages
63 - 70
Database
ISI
SICI code
0003-2670(20010801)440:1<63:RODOCI>2.0.ZU;2-U
Abstract
A fast screening procedure for water samples contaminated with chlorobenzen e was developed. The volatile organic compounds were purged from water samp les and their concentration was determined with ion mobility spectrometry ( IMS) using corona discharge (CD) ionization. This analytical procedure perm its the determination of chlorobenzene with a concentration range of 3-30 m g/l. The analytical results are available within 5 min. In order to test th e general suitability of the method developed, the influence of sample matr ix, especially of accompanying chlorinated aromatics, on ion mobility spect ra of chlorobenzene was investigated. (C) 2001 Elsevier Science B.V. All ri ghts reserved.