F. Mouret et al., In-vitro measurements of the regurgitation of mechanical mitral heart valve prostheses in case of atrial fibrillation, J HEART V D, 10(2), 2001, pp. 264-268
Background and aim of the study: The characterization of heart valve prosth
eses requires regurgitation tests to be conducted in pulsatile flow. Althou
gh tests are generally conducted in accordance with hemodynamic conditions
of a healthy young man, heart valve implantation is often associated with o
ther pathology, such as atrial fibrillation. To run more realistic trials,
four mechanical heart valve prostheses were tested with, and then without,
atrial contraction.
Methods: The dual activation simulator (DAS) allow simulation of physiologi
cally normal and pathological flows through the mitral valve. The DAS compr
ises silicon-based cavities, is activated by pumps, and was equipped succes
sively with monoleaflet (Bjork-Shiley, Medtronic Hall) and bileaflet (St. J
ude Medical, CarboMedics) valves. Each valve (mitral, size 27 mm) was teste
d under two conditions (with and without atrial contraction) at a mean flow
rate of 3 l/min of glycerol/water solution (analog blood viscosity).
Results: Leakage volumes were of the same magnitude as the precision of the
instruments. Respectively, closing volumes increased from normal condition
s to atrial fibrillation from 3.2 to 5.1 ml for Bjork-Shiley, from 4.6 to 6
.3 ml for Medtronic Hall, from 5 to 6.6 ml for St. Jude Medical, and from 5
.2 to 5.4 ml for CarboMedics. The standard deviation was below the precisio
n of measurements (+/-0.5 ml).
Conclusion: Without atrial contraction, the valves seemed to be closed by b
ackward flow only, thus confirming earlier reports. This study showed that
different heart valves behave differently in pathological situations with r
egard to their design; this must be considered when selecting a valve for i
mplantation.