M. Alecci et al., Radio frequency magnetic field mapping of a 3 tesla birdcage coil: Experimental and theoretical dependence on sample properties, MAGN RES M, 46(2), 2001, pp. 379-385
Citations number
35
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
The IRF B-1 distribution was studied, theoretically and experimentally, in
phantoms and in the head of volunteers using a 3 T MRI system equipped with
a birdcage coil. Agreement between numerical simulation and experiment dem
onstrates that B-1 distortion at high field can be explained with 3D full-M
axwell calculations. It was found that the B-1 distribution in the transver
se plane is strongly dependent on the dielectric properties of the sample.
We show that this is a consequence of RF penetration effects combined with
RF standing wave effects. In contrast, along the birdcage coil z-axis the B
-1 distribution is determined mainly by the coil geometry. In the transvers
e plane, the region of B-1 uniformity (within 10% of the maximum) was 15 cm
with oil, 6 cm with distilled water, 11 cm with saline, and 10 cm in the h
ead. Along z the B-1 uniformity was 9 cm with phantoms and 7 cm in the head
. (C) 2001 Wiley-Liss, Inc.