Radio frequency magnetic field mapping of a 3 tesla birdcage coil: Experimental and theoretical dependence on sample properties

Citation
M. Alecci et al., Radio frequency magnetic field mapping of a 3 tesla birdcage coil: Experimental and theoretical dependence on sample properties, MAGN RES M, 46(2), 2001, pp. 379-385
Citations number
35
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
MAGNETIC RESONANCE IN MEDICINE
ISSN journal
07403194 → ACNP
Volume
46
Issue
2
Year of publication
2001
Pages
379 - 385
Database
ISI
SICI code
0740-3194(200108)46:2<379:RFMFMO>2.0.ZU;2-H
Abstract
The IRF B-1 distribution was studied, theoretically and experimentally, in phantoms and in the head of volunteers using a 3 T MRI system equipped with a birdcage coil. Agreement between numerical simulation and experiment dem onstrates that B-1 distortion at high field can be explained with 3D full-M axwell calculations. It was found that the B-1 distribution in the transver se plane is strongly dependent on the dielectric properties of the sample. We show that this is a consequence of RF penetration effects combined with RF standing wave effects. In contrast, along the birdcage coil z-axis the B -1 distribution is determined mainly by the coil geometry. In the transvers e plane, the region of B-1 uniformity (within 10% of the maximum) was 15 cm with oil, 6 cm with distilled water, 11 cm with saline, and 10 cm in the h ead. Along z the B-1 uniformity was 9 cm with phantoms and 7 cm in the head . (C) 2001 Wiley-Liss, Inc.