Sa. Naqvi et al., Analysing collimator structure effects in head-scatter calculations for IMRT class fields using scatter raytracing, PHYS MED BI, 46(7), 2001, pp. 2009-2028
The frequent blocking of the irradiated volume in intensity modulated radia
tion therapy (IMRT) makes the head-scatter fraction of the incident photon
fluence more significant than that in conventional therapy with open fields
. On the other hand, certain collimator configurations block scatter photon
s directed to a given observation point while allowing primary photons to b
e transmitted. The 'anomalous blocking' makes the primary field a poor indi
cator of the scatter fluence. Since large MU-to-cGy ratios in IMRT can magn
ify head-scatter uncertainties, it becomes necessary to accurately model bo
th the effective scatter source and the collimator structure that lit-nits
the scatter reaching the irradiated volume. First we obtain a dual-source m
odel, using a Taylor series expansion to derive the effective scatter sourc
e distribution from the data measured for the Elekta SL20 linac equipped wi
th a multi-leaf collimator (MLC). Then, using a raytracing algorithm, we ca
lculate the transmission of scatter rays from the effective scatter source
plane to points in the patient plane. The method can account for the anomal
ous blocking of scatter by the MLC leaves and the backup diaphragms. For a
variety of collimator settings tested, the calculations agree with measurem
ents to an accuracy of 0.002 Psi (10 x 10) where Psi (10 x 10) is the total
(primary + scatter) photon fluence of an open 10 x 10 cm(2) field for the
same MU delivered. Although the significance of collimator structure in IMR
T depends strongly on fields shapes employed for the delivery, potential Cu
mulative errors on the order of a few per cent can be avoided in fluence ca
lculations if the proposed method is used.