J. Perez-calatayud et al., A Monte Carlo study of dose rate distribution around the specially asymmetric CSM3-a Cs-137 source, PHYS MED BI, 46(7), 2001, pp. N169-N174
The CSM3 Cs-137 type stainless-steel encapsulated source is widely used in
manually afterloaded low close rate brachytherapy. A specially asymmetric s
ource. CSM3-a, has been designed by CIS Bio International (France) substitu
ting the eyelet side seed with an inactive material in the CSM3 source. Thi
s modification has been done in order to allow a uniform dose level over th
e upper vaginal surface when this 'linear' source is inserted at the top of
the dome vaginal applicators. In this study the Monte Carlo GEANT3 simulat
ion code. incorporating the source geometry in detail, was used to investig
ate the dosimetric characteristics of this special CSM3-a Cs-137 brachyther
apy source. The absolute dose rate distribution in water around this source
was calculated and is presented in the form of an along-away table. Compar
ison of Sievert integral type calculations with Monte Carlo results are dis
cussed.