C. Yang et al., Phase-referenced interferometer with subwavelength and subhertz sensitivity applied to the study of cell membrane dynamics, OPTICS LETT, 26(16), 2001, pp. 1271-1273
We report a highly sensitive means of measuring cellular dynamics with a no
vel interferometer that can measure motional phase changes. The system is b
ased on a modified Michelson interferometer with a composite laser beam of
1550-nm low-coherence light and 775-nm CW light. The sample is prepared on
a coverslip that is highly reflective at 775 run. By referencing the hetero
dyne phase of the 1550-nm light reflected from the sample to that of the 77
5-nm light reflected from the coverslip, small motions in the sample are de
tected, and motional artifacts from vibrations in the interferometer are co
mpletely eliminated. We demonstrate that the system is sensitive to motions
as small as 3.6 nm and velocities as small as 1 nm/s. Using the instrument
, we study transient volume changes of a few (approximately three) cells in
a monolayer immersed in weakly hypotonic and hypertonic solutions. (C) 200
1 Optical Society of America.