El. Chan et al., COMPARISON OF THE SANOFI DIAGNOSTICS PASTEUR CHLAMYDIA MICROPLATE EIASHORTENED ASSAY WITH THE ORIGINAL STANDARD ASSAY AND CELL-CULTURE, Journal of Medical Microbiology, 46(8), 1997, pp. 711-713
The new Sanofi Diagnostics Pasteur Chlamydia Microplate EIA shortened
assay was evaluated- by comparison with the original standard assay an
d cell culture, A total of 853 paired male and female genital tract sp
ecimens was tested with both Sanofi Chlamydia Microplate EIA shortened
and standard assays and the results were compared with those of cell
culture, For confirmation, a blocking assay run in the shortened forma
t was used, Discrepancies between the three methods were resolved by a
direct fluorescent antibody (DFA) test on the EIA samples or the cult
ure retentate, or both, After resolution of discrepant results, the st
andard assay had a sensitivity, specificity, positive predictive value
and negative predictive value of 98.5%, 100%, 100% and 99.9%, respect
ively, The shortened assay results were 100%, 100%, 100% and 100%, res
pectively, The shortened assay takes approximately 1.5 h less time tha
n the standard assay and this study demonstrated that they have equiva
lent sensitivity and specificity. The improvement in turnaround time e
nables results to be reported on the same day.