Washing machine associated 50 hz detected as ventricular fibrillation by an implanted cardioverter defibrillator

Citation
X. Sabate et al., Washing machine associated 50 hz detected as ventricular fibrillation by an implanted cardioverter defibrillator, PACE, 24(8), 2001, pp. 1281-1283
Citations number
4
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
Journal title
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
ISSN journal
01478389 → ACNP
Volume
24
Issue
8
Year of publication
2001
Pages
1281 - 1283
Database
ISI
SICI code
0147-8389(200108)24:8<1281:WMA5HD>2.0.ZU;2-P
Abstract
This case report describes a patient with an automatic ICD who suffered a d efibrillation shock without warning symptoms. An electrical interference ca n be observed in the stored EGM of the episode. The patient explained that the moment he felt the shock he was touching a washing machine, After corre ct grounding of this machine the patient did not suffer more inappropriate shocks.