Delayed restoration of atrioventricular synchrony with beat-to-beat mode switch

Citation
F. Bode et al., Delayed restoration of atrioventricular synchrony with beat-to-beat mode switch, PACE, 24(8), 2001, pp. 1286-1288
Citations number
7
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
Journal title
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
ISSN journal
01478389 → ACNP
Volume
24
Issue
8
Year of publication
2001
Pages
1286 - 1288
Database
ISI
SICI code
0147-8389(200108)24:8<1286:DROASW>2.0.ZU;2-1
Abstract
This case report describes a patient with complete AV block and a VDD pacem aker who experienced repetitive episodes of symptomatic bradycardia. Episod es occurred due to activation of an automatic beat-to-beat mode switch algo rithm. After mode switch to VDI operation, the pacemaker failed to immediat ely switch back to AV synchronous pacing when regular sinus rhythm (greater than or equal to 100 beats per minute) resumed despite adequate P wave sen sing. Dizziness was felt for up to several minutes of asynchronous pacing a t the lower rate limit until VDD mode was restored. Episodes were completel y eliminated by programming the mode switch function from an automatic beat -to-beat algorithm to a fixed rate algorithm.