This case report describes a patient with complete AV block and a VDD pacem
aker who experienced repetitive episodes of symptomatic bradycardia. Episod
es occurred due to activation of an automatic beat-to-beat mode switch algo
rithm. After mode switch to VDI operation, the pacemaker failed to immediat
ely switch back to AV synchronous pacing when regular sinus rhythm (greater
than or equal to 100 beats per minute) resumed despite adequate P wave sen
sing. Dizziness was felt for up to several minutes of asynchronous pacing a
t the lower rate limit until VDD mode was restored. Episodes were completel
y eliminated by programming the mode switch function from an automatic beat
-to-beat algorithm to a fixed rate algorithm.