Lead configuration for defibrillator implantation in a patient with congenital heart disease and a mechanical prosthetic tricuspid valve

Citation
Ct. Leng et al., Lead configuration for defibrillator implantation in a patient with congenital heart disease and a mechanical prosthetic tricuspid valve, PACE, 24(8), 2001, pp. 1291-1292
Citations number
3
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
Journal title
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
ISSN journal
01478389 → ACNP
Volume
24
Issue
8
Year of publication
2001
Pages
1291 - 1292
Database
ISI
SICI code
0147-8389(200108)24:8<1291:LCFDII>2.0.ZU;2-I
Abstract
The authors devised a nonthoracotomy defibrillation system for a patient wi th a prosthetic tricuspid valve using existing technology and previously es tablished implantation techniques. Their lead configuration deviates substa ntially from existing designs in its primary use of a coronary sinus defibr illation coil and a left-sided subcutaneous array to distribute current acr oss the ventricular myocardium.