An integrated electrochemical capillary liquid chromatography-dual microelectrode system for bromine based reaction detection

Citation
Hp. Zhou et al., An integrated electrochemical capillary liquid chromatography-dual microelectrode system for bromine based reaction detection, ANALYST, 126(8), 2001, pp. 1252-1256
Citations number
24
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYST
ISSN journal
00032654 → ACNP
Volume
126
Issue
8
Year of publication
2001
Pages
1252 - 1256
Database
ISI
SICI code
0003-2654(200108)126:8<1252:AIECLC>2.0.ZU;2-B
Abstract
Amperometric reaction detection of model peptides containing either a thiol or disulfide was successfully achieved with a novel coupled capillary liqu id chromatography-dual microelectrode system. Detection is facilitated via dual platinum wire electrodes fabricated directly on a liquid chromatograph y capillary column. The detection strategy is based on the electrogeneratio n of bromine at the upstream working electrode followed by reduction of the bromine to bromide at the downstream working electrode. When bromine react s with analyte eluting from the capillary column. the amount of bromine tha t reaches the downstream electrode decreases. As a result. the current at t he downstream electrode will decrease in proportion to the amount of analyt e eluting from the capillary column. Glutathione., glutathione disulfide, a lpha -TGF and oxytocin were used to characterize the system and to determin e system parameters. The feasibility of the determination of disulfide cont aining peptides is demonstrated with glutathione disulfide, alpha -TGF and oxytocin.