Gq. Chen et al., Studies on high-resolution hadamard transform microscopic image detection system using one-dimensional mechanical mask, CHEM J CH U, 22(7), 2001, pp. 1114-1116
A novel Hadamard transform(HT) microscopic image detection system which can
generate 511 x 512 pixel format images of fluorescence or absorptive light
for small samples has been developed, when spatial multiplexing is perform
ed with a 511-slit one-dimensional Hadamard mask and a highly sensitive lin
ear CCD. The weak native fluorescence imaging for Zephyranthes candida(Lind
l.) Herb. Pollen cells can be realized within 1 min. The spatial resolution
of 0.6 mum per pixel has been achieved with a 40X objective. The imaging s
peed, sensitivity and spatial resolution are satisfactory, hence the system
has a good application prospect in cell biology and medicine. Some factors
that influence imaging speed and quality have been discussed preliminarily
.