Mv. Narayanan et al., Optimization of regularization of attenuation and scatter-corrected Tc-99mcardiac SPECT studies for defect detection using hybrid images, IEEE NUCL S, 48(3), 2001, pp. 785-789
Through means of a receiver-operating-characteristics study, we optimize th
e iteration number and three-dimensional (3-D) Gaussian postfiltering of Tc
-99m cardiac emission ordered-subset expectation-maximization (OSEM) recons
tructions that implement corrections for both attenuation and scatter. Hybr
id images, wherein artificial perfusion defects were added to clinical pati
ent studies that were read as being normally perfused, were used for this o
ptimization. The test conditions included three different iteration numbers
of OSEM (1, 5, and 10) using four angles per subset, followed by 3-D Gauss
ian low-pass filtering at each iteration level. The level of Gaussian low-p
ass filtering was varied using standard deviations (a) of 0.6, 0.75, 1, and
1.25 pixels, in addition to a case where no postfiltering was applied. Fou
r observers read 80 images for each of the 15 test conditions being investi
gated, providing confidence ratings as to the presence or absence of perfus
ion defects. Results indicate that at all iterations, optimum detection per
formance is obtained for a broad plateau or range of postfilters (sigma = 0
.6 to 1 pixel). As expected, a gradual reduction in performance is seen at
either end of this broad maximum where the images either have been very hea
vily smoothed or have very little postfiltering. Finally, one iteration of
OSEM appears to be the appropriate choice since no significant improvement
in detection accuracy was observed with increasing iteration number as long
as the reconstructions are postfiltered with or in the range of 0.6 to I p
ixel.