Optimization of dedicated scintimammography procedure using detector prototypes and compressible phantoms

Citation
S. Majewski et al., Optimization of dedicated scintimammography procedure using detector prototypes and compressible phantoms, IEEE NUCL S, 48(3), 2001, pp. 822-829
Citations number
24
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science","Nuclear Emgineering
Journal title
IEEE TRANSACTIONS ON NUCLEAR SCIENCE
ISSN journal
00189499 → ACNP
Volume
48
Issue
3
Year of publication
2001
Part
2
Pages
822 - 829
Database
ISI
SICI code
0018-9499(200106)48:3<822:OODSPU>2.0.ZU;2-A
Abstract
Results are presented on the optimization of the design and use of dedicate d compact scintimammography gamma cameras. Prototype imagers with a field o f view (FOV) of 5 x 5 cm(2), 10 x 10 cm(2), and 15 x 20 cm(2) were used in either a dual modality mode as an adjunct technique to digital X-ray mammog raphy imagers or as stand-alone instruments such as dedicated breast SPECT and planar imagers. Experimental data were acquired to select the best imag ing modality (SPECT or planar) to detect small lesions using Tc-99m radio-l abeled pharmaceuticals. In addition, studies were performed to optimize the imaging geometry. Results suggest that the preferred imaging geometry is p lanar imaging with two opposing detector heads while the breast is under co mpression. However, further study of the dedicated breast SPECT is warrante d.