Ee. Klein et Da. Low, Interleaf leakage for 5 and 10 mm dynamic multileaf collimation systems incorporating patient motion, MED PHYS, 28(8), 2001, pp. 1703-1710
Citations number
21
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Use of dynamic multileaf collimation (DMLC) for intensity modulated radiati
on therapy (IMRT) is accelerating. Delivery systems have the ailment of int
erleaf leakage (IL). This is compounded by the inefficiency of IMRT deliver
y, estimated to be a factor of 5 for DMLC. With IL on the order of 4%, it i
s possible to deliver as much as 20% of the prescribed dose to nonprescribe
d regions. However, IL is characterized by narrow Gaussian peaks of similar
to0.5-1.0 mm full-width-half-maximum (FWHM). We performed a leakage study
for 5 and 10 mm leaf systems, accounting for intratreatment and intertreatm
ent motions. In solid phantoms, film was placed perpendicular to beams. DML
C patterns delivered step-wedged distributions. The same field was duplicat
ed using a collimating jaw in a segmented fashion to obtain baseline data o
f primary and scatter contributions. Longitudinal shifts up to 4 mm and ang
ulations up to 4 degrees were introduced during beam delivery by running mu
ltiple patterns, to arrive at a composite delivery. The intent of these rig
id body motion experiments was to replicate patient motion. Clinical IMRT f
ields using segmented MLC were also tested. Films were scanned and converte
d to dose. A microionization chamber confirmed film data at discrete points
. In all cases shifts diminished IL peak values. In the step-wedge case, th
e net 18 MV IL peaks diminished from 3.6% to 3.2% for the 10 mm system. The
5 mm. system IL values decreased from 4.0% to 3.2% with a 2 mm shift but i
ncreased to 4.0% with 4 nun shifts. The clinical field data followed the sa
me pattern with a washing out of peak values, but the overall transmission
to shielded regions slightly increased. Therefore nonprescribed regions are
influenced by an effective transmission value rather than discrete peak IL
values. The 5 mm leaf system does not introduce increased IL and is an app
ropriate system for IMRT. (C) 2001 American Association of Physicists in Me
dicine.