Light shift in an optically pumped caesium-beam frequency standard

Citation
Jw. Jun et al., Light shift in an optically pumped caesium-beam frequency standard, METROLOGIA, 38(3), 2001, pp. 221-227
Citations number
12
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
METROLOGIA
ISSN journal
00261394 → ACNP
Volume
38
Issue
3
Year of publication
2001
Pages
221 - 227
Database
ISI
SICI code
0026-1394(2001)38:3<221:LSIAOP>2.0.ZU;2-I
Abstract
An evaluation of the light shift in an optically pumped caesium-beam freque ncy standard is presented. We show that the light shift can be accurately e stimated by taking into account the angular anisotropy of the spontaneous p hoton emission, the neglect of which is shown to give rise to an error comp arable to the value of the shift. Our data refer to a complete set of optic al transitions for the caesium. D-2 line relevant to the KRISS-1 caesium-be am frequency standard.