Response of CZT drift-strip detector to X- and gamma rays

Citation
I. Kuvvetli et al., Response of CZT drift-strip detector to X- and gamma rays, RADIAT PH C, 61(3-6), 2001, pp. 457-460
Citations number
3
Categorie Soggetti
Physics
Journal title
RADIATION PHYSICS AND CHEMISTRY
ISSN journal
0969806X → ACNP
Volume
61
Issue
3-6
Year of publication
2001
Pages
457 - 460
Database
ISI
SICI code
0969-806X(200106)61:3-6<457:ROCDDT>2.0.ZU;2-D
Abstract
The drift-strip method for improving the energy response of a CdZnTe (CZT) detector to hard X- and gamma rays is discussed. Results for a 10 x 10 x 3 mm(3) detector crystal demonstrate a remarkable improvement of the energy r esolution. The full width at half maximum (FWHM) is 2.18 keV (3.6%), 2.45 k eV (3.0%), 2.86 keV (0.8 %) and 3.89 keV (0.6%) at 60, 80, 356 and 661 keV, respectively. The resolution is limited by electronic noise below 100 keV. (C) 2001 Elsevier Science Ltd. All rights reserved.