The effect of humidity on the accuracy of measurement of an electret radondosimeter

Citation
Rh. Mahat et al., The effect of humidity on the accuracy of measurement of an electret radondosimeter, RADIAT PH C, 61(3-6), 2001, pp. 489-490
Citations number
1
Categorie Soggetti
Physics
Journal title
RADIATION PHYSICS AND CHEMISTRY
ISSN journal
0969806X → ACNP
Volume
61
Issue
3-6
Year of publication
2001
Pages
489 - 490
Database
ISI
SICI code
0969-806X(200106)61:3-6<489:TEOHOT>2.0.ZU;2-W
Abstract
Test has been conducted for the response of electret radon dosimeters with varying levels of humidity. At a constant value of radon concentration, mea sured values have been found to increase linearly with increasing humidity, from 30% RH up to a value of similar to 85% RH. On this basis we argue for the establishment of a humidity-dependent calibration factor. (C) 2001 Els evier Science Ltd. All rights reserved.