Ym. Huang et Zj. Zhang, Flow injection chemiluminescent analysis of isoniazid by direct hexacyanoferrate(III) oxidation, ANAL LETTER, 34(10), 2001, pp. 1703-1710
A very simple chemiluminescence (CL) flow system is described for the deter
mination of isoniazid based on its direct oxidation by hexacyanoferrate(III
) solution in an alkaline media. The system responds linearly to isoniazid
concentration in the range 0.05-2 mug/mL with a detection limit (3 sigma) o
f 0.01 mug/mL. Relative standard deviation (RSD) is 1.6% for 0.5 mug/mL of
isoniazid (n=11). The system has been successfully applied to the determina
tion of isoniazid in pharmaceutical preparations. The measurement could be
performed in 30 s with sampling rate of 120 times per hour.