Better-resolution measurement of vertical scattering angle in a new ion-optical mode of spectrometer "Grand Raiden"

Citation
H. Fujita et al., Better-resolution measurement of vertical scattering angle in a new ion-optical mode of spectrometer "Grand Raiden", NUCL INST A, 469(1), 2001, pp. 55-62
Citations number
13
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
469
Issue
1
Year of publication
2001
Pages
55 - 62
Database
ISI
SICI code
0168-9002(20010801)469:1<55:BMOVSA>2.0.ZU;2-H
Abstract
Demand for near-zero-degree measurements using a magnetic spectrometer is i ncreasing with the growing interest in the study of L = 0 nuclear excitatio n modes. For the precise determination of scattering angles near 0 degrees, both horizontal and vertical scattering angle components have to be measur ed with good accuracy. It is, however, not easy to realize a good vertical angle resolution using modern high-resolution magnetic spectrometers with s mall vertical angle magnifications, like "Grand Raiden" at RCNP or "K600" a t IUCF. A new ion-optical mode which enables the precise determination of vertical components of scattering angles from vertical position measurements in the focal plane has been developed for these spectrometers. For the wide range of spatial as well as momentum acceptances, vertical angle resolution bette r than 8 mrad is realized for Grand Raiden at RCNP. (C) 2001 Elsevier Scien ce B.V. All rights reserved.