Quantitative measurement of soft-X-ray spectrum using transmission gratingspectrometer

Citation
Jm. Yang et al., Quantitative measurement of soft-X-ray spectrum using transmission gratingspectrometer, ACT PHY C E, 50(9), 2001, pp. 1723-1728
Citations number
15
Categorie Soggetti
Physics
Journal title
ACTA PHYSICA SINICA
ISSN journal
10003290 → ACNP
Volume
50
Issue
9
Year of publication
2001
Pages
1723 - 1728
Database
ISI
SICI code
1000-3290(200109)50:9<1723:QMOSSU>2.0.ZU;2-R
Abstract
Transmission grating efficiencies for soft X-ray and quantum efficiencies o f X-ray CCD in the energy range of 100-1500eV were carefully calibrated on Beijing Synchrotron Radiation Facility. A model is developed to simulate th e calibrated grating efficiencies, another model is developed to calculate the CCD quantum efficiencies. The effect of CCD surface contamination on CC D quantum efficiency is also studied. Finally, a method for unfolding the s oft X-ray spectrum measured with transmission grating spectrometer is intro duced, in which the grating efficiencies, CCD quantum efficiencies and grat ing higher order overlapping have been taken into account. The developed qu antitative measurement method has been applied to obtain X-ray spectrum fro m gold plasma.