Rm. Lindsay et al., Relationship between effective ionic dialysance and in vivo urea clearanceduring hemodialysis, AM J KIDNEY, 38(3), 2001, pp. 565-574
Effective ionic dialysance (EID) can be measured from dialyzer inlet and ou
tlet conductivity changes following two steps of dialysate conductivity. Re
lationships between EID and In vivo urea clearances were studied four times
per hemodialysis treatment in eight patients, each undergoing six hemodial
ysis treatments (192 data sets). Dialyzer blood flow was varied from 190 to
500 mL/min. Dialysate flow was constant (751 to 771 mL/min), and a standar
d dialyzer (700 HG; Cobe, Lakewood, CO) was used. Double samples were drawn
for arterial, venous, and dialysate urea measurements. Two laboratory valu
es were missing. Twelve unreliable laboratory values indicated by divergent
results were excluded. Urea clearances were calculated by formulae convert
ing whole-blood to blood-water urea clearances. EID was measured using Dias
can (Gambro-Dasco, Medolla, Italy). Mass balance was checked by comparison
of dialysate and blood-water urea clearances. Divergent results between dia
lysate and blood-water urea clearance values led to the exclusion of an add
itional three laboratory values. A small error (4.2%) in urea mass balance
was found (dialysate greater than blood-water urea clearances). A total of
175 data sets were compared. EID showed excellent correlation with blood-wa
ter urea clearances (r = 0.92) over the line of identity, with a mean diffe
rence of -3.5 mL/min (-1%), and similarly with dialysate urea clearances (r
= 0.92; mean difference, -13.4 mL/min; -5%). For both blood- and dialysate
-side comparisons, differences Increased with greater clearances. Because E
ID is an effective clearance and urea clearance is a measure of dialyzer cl
earance, the curves were corrected for cardiopulmonary recirculation; acces
s recirculation was zero (Transonic monitor; Transonic Systems Inc, Ithaca,
NY). For cardiopulmonary recirculation correction, cardiac output and acce
ss flows were assumed to be 6.4 L and 1.46 L/min. Corrected data show EID c
orrelates with blood-side urea clearance (r = 0.92), with a mean difference
of +7.3 mL/min (3.3%), and is constant over the range of clearances. EID c
orrelated with dialysate urea clearance (r = 0.92) with virtually no differ
ence. The difference on the blood side is consistent with the urea mass bal
ance error found. These data indicate that EID using Diascan can provide an
accurate indication of effective urea clearances obtained during hemodialy
sis and is of value in monitoring dialysis adequacy. (C) 2001 by the Nation
al Kidney Foundation, Inc.