The aim of the study is to improve dose uniformity in the body by the compe
nsator-rice and to decrease the dose to the lung by the partial lung block.
Rando phantom supine was set up to treat bilateral fields with a 15 MV lin
ear accelerator at 415 cm treatment distance. The experimental procedure in
cluded three parts. The first part was the bilateral irradiation without ri
ce compensator, and the second part was with rice compensator. In the third
part, rice compensator and partial lung block were both used. The results
of thermoluminescent dosimeters measurements indicated that without rice co
mpensator the dose was non-uniform. Contrarily, the average dose homogeneit
y with rice compensator was measured within +/-5%, except for the thorax re
gion. Partial lung block can reduce the dose which the lung received. This
is a simple method to improve the dose homogeneity and to reduce the lung d
ose received. The compensator-rice is cheap, and acrylic boxes are easy to
obtain. Therefore, this technique is suitable for more studies. (C) 2001 El
sevier Science Ltd. All rights reserved.