Dose compensation of the total body irradiation therapy

Citation
Jp. Lin et al., Dose compensation of the total body irradiation therapy, APPL RAD IS, 55(5), 2001, pp. 623-630
Citations number
17
Categorie Soggetti
Multidisciplinary
Journal title
APPLIED RADIATION AND ISOTOPES
ISSN journal
09698043 → ACNP
Volume
55
Issue
5
Year of publication
2001
Pages
623 - 630
Database
ISI
SICI code
0969-8043(200111)55:5<623:DCOTTB>2.0.ZU;2-S
Abstract
The aim of the study is to improve dose uniformity in the body by the compe nsator-rice and to decrease the dose to the lung by the partial lung block. Rando phantom supine was set up to treat bilateral fields with a 15 MV lin ear accelerator at 415 cm treatment distance. The experimental procedure in cluded three parts. The first part was the bilateral irradiation without ri ce compensator, and the second part was with rice compensator. In the third part, rice compensator and partial lung block were both used. The results of thermoluminescent dosimeters measurements indicated that without rice co mpensator the dose was non-uniform. Contrarily, the average dose homogeneit y with rice compensator was measured within +/-5%, except for the thorax re gion. Partial lung block can reduce the dose which the lung received. This is a simple method to improve the dose homogeneity and to reduce the lung d ose received. The compensator-rice is cheap, and acrylic boxes are easy to obtain. Therefore, this technique is suitable for more studies. (C) 2001 El sevier Science Ltd. All rights reserved.