Improvement in the sample mass-change-method used for determination of thebulk etch rate of CR-39 neutron detector

Citation
Ia. Matiullah,"tahiri et al., Improvement in the sample mass-change-method used for determination of thebulk etch rate of CR-39 neutron detector, JPN J A P 1, 40(5A), 2001, pp. 3474-3475
Citations number
14
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science
Volume
40
Issue
5A
Year of publication
2001
Pages
3474 - 3475
Database
ISI
Abstract
The bulk etch rate is an important parameter of any etch-track detector. An accurate measurement of the bulk etch rate is, therefore, highly desirable in the calibration of a CR-39-based neutron dosimeter. Amongst several ava ilable bulk etch rate measuring techniques, the sample mass change techniqu e is claimed to be the most accurate one. We carried out systematic experim ents and found that even this method has some limitations. The problems ass ociated with this method are discussed. To circumvent these problems, modif ications in the sample mass change technique have been suggested.