Ultrahigh scanning speed optical coherence tomography using optical frequency comb generators

Citation
Sj. Lee et al., Ultrahigh scanning speed optical coherence tomography using optical frequency comb generators, JPN J A P 2, 40(8B), 2001, pp. L878-L880
Citations number
20
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science
Volume
40
Issue
8B
Year of publication
2001
Pages
L878 - L880
Database
ISI
SICI code
Abstract
A novel optical coherence tomography system without any moving part.,, for depth scanning was proposed and demonstrated. It was accomplished using two optical frequency comb generators, as broad-band spectrum generators, inst ead of moving part,, for depth scanning. A high scanning speed of more than 12,5 km/s and also a high repetition rate of 500 kHz were achieved in a lo ng scan range of 25 mm. The resolution was about 100 mum.