R. Gonzalez-hernandez et al., Reversed phase high performance liquid chromatographic determination of cefixime in bulk drugs, J LIQ CHR R, 24(15), 2001, pp. 2315-2324
A new technique for the quantitative determination of cefixime trihydrate i
n bulk drugs by high performance liquid chromatography was developed using
simple reagents. Finally, chosen conditions of analysis were as follows: Li
Chrospher 100 RP - 18 (250 x 4 mm I.D.) column, mobile phase consisting of
phosphate buffer pH 7.0 and acetonitrile (93:7, v/v), flow rate of 0,8 mL/m
in, loop of 20 muL and UV detection at 287 nm. The prospective validation o
f this technique showed that it is linear in a range of 0.1 to 0.6 mg/mL (r
= 0.9997), sensitive (0.3 %), precise (within-a-day repeatability, RSD = 1
.0 %, day-to-day repeatability RSD = 1.3 %), accurate and selective (cefixi
me can be determined in presence of its related compounds). The limits of d
etection and quantitation are 37 ng (0.3 %) and 128 ng (1.1 %), respectivel
y, relative to a 0.6 mg/mL solution.