Blade-type X-ray beam position monitors for SPring-8 undulator beamlines

Citation
H. Aoyagi et al., Blade-type X-ray beam position monitors for SPring-8 undulator beamlines, NUCL INST A, 467, 2001, pp. 252-255
Citations number
5
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
467
Year of publication
2001
Part
1
Pages
252 - 255
Database
ISI
SICI code
0168-9002(20010721)467:<252:BXBPMF>2.0.ZU;2-J
Abstract
The X-ray beam position monitors had been designed and installed for SPring -8 insertion device beamlines. These monitors are being utilized for photon beam diagnostics. The beam from the standard undulator in SPring-8 has the total power of 11 kW and the power density of 470 kW/mrad(2), typically. E ach monitor has four CVD diamond blades coated with metal for detector head s. We have already introduced three styles of monitors to match various ins ertion devices in SPring-8. A standard style, or a fixed-blade style, is us ed mainly for a standard in-vacuum undulator beamlines. A horizontal-blade- drive style and a four-blade-drive style are used for beamlines of a wiggle r and a twin helical undulator that have wide power distributions, and for figure-8 undulators that have asymmetric power distributions, respectively. This report describes the design and the structure of these monitors and t he beam-tests for the photon beam diagnostics in detail. (C) 2001 Published by Elsevier Science B.V.