High energy synchrotron X-ray focusing by fixed exit bender
Citation
Y. Yoneda et al., High energy synchrotron X-ray focusing by fixed exit bender, NUCL INST A, 467, 2001, pp. 370-372
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
SICI code
0168-9002(20010721)467:<370:HESXFB>2.0.ZU;2-6
Abstract
The performance of sagittal focusing for hard X-rays with a cylindrical ben
t crystal at SPring-8 is described. This bending mechanism is designed for
the SPring-8 standard double crystal monochromator of bending-magnet beamli
nes. Therefore, a fixed height exit for a wide energy range (5 - 150 keV) i
s required. The results underline that focusing in a wide energy band was a
chieved by combining an adjustable inclined double crystal monochromator an
d a fixed height exit bending mechanism. The monochromator was maintained a
t a constant focus at the sample position by optimizing the bent crystal in
the energy range from 60 to 150 keV. (C) 2001 Elsevier Science B.V. All ri
ghts reserved.