Diamond double-crystal monochromator in Bragg geometry installed on BL-11XU at SPring-8

Citation
M. Marushita et al., Diamond double-crystal monochromator in Bragg geometry installed on BL-11XU at SPring-8, NUCL INST A, 467, 2001, pp. 392-395
Citations number
6
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
467
Year of publication
2001
Part
1
Pages
392 - 395
Database
ISI
SICI code
0168-9002(20010721)467:<392:DDMIBG>2.0.ZU;2-8
Abstract
We present here the feature of the diamond double-crystal monochromator in Bragg geometry installed on a standard undulator beamline at SPring-8. The crystal was manufactured by Sumitomo Electric Industries, Ltd., whose size was 8.6 mm (w) x 3.5 mm. (l) x 0.35 mm. (t) for the first crystal and 10 mm . (w) x 4.7 mm (l) x 0.39 mm (t) for the second. The feature of the monochr omator was tested by rocking curve measurements as a function of the total power and of the energy that impinged on the crystal. As a result, no signi ficant increase of the full-width at half-maximum was observed up to a tota l power of 330 W on the first crystal. We discuss the experimental results with the comparison to the calculated FWHM with use of the beamline paramet ers. (C) 2001 Elsevier Science B.V. All rights reserved.