Source size conserving broad band monochromators of fixed exit geometry for high energy synchrotron radiation

Citation
U. Lienert et al., Source size conserving broad band monochromators of fixed exit geometry for high energy synchrotron radiation, NUCL INST A, 467, 2001, pp. 659-662
Citations number
7
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
467
Year of publication
2001
Part
1
Pages
659 - 662
Database
ISI
SICI code
0168-9002(20010721)467:<659:SSCBBM>2.0.ZU;2-0
Abstract
The degradation of the effective source size by energy tunable, fixed exit, broad band monochromators for high energy synchrotron radiation (40-100 ke V) is measured to be below 2 mu rad. Two monochromator schemes consisting o f either two bent Lane crystals or two SiGe gradient crystals in a non-disp ersive setting are characterized. The relative energy bandwidth is about 0. 1% and reflectivities are above 80%. The source size broadening and band wi dths are found to be in agreement with predictions by a kinematical model. (C) 2001 Published by Elsevier Science B.V.