The 4W1B is a X-ray monochromator beamline for XAFS at BSRF. During the upg
rading phase, we have redesigned the monochromator to improve the performan
ce of the beamline. It is a goniometer based, fixed exit double crystal mon
ochromator. A mechanical linkage is employed to adjust the distance between
the surfaces of the two crystals as the Bragg angle is changed to keep the
outgoing beam direction constant. The whole mechanism is driven by only on
e stepping motor. The testing result shows that over the scanning range of
5-30 degrees, the shift of outgoing beam position is less then 70 mum in th
e vertical direction. The basic principle, the mechanical realization, and
the error analysis are discussed in detail. The performance and the testing
results are also presented in this paper. (C) 2001 Elsevier Science B.V. A
ll rights reserved.