Assessment of the linear reference air kerma rate of Ir-192 wires

Citation
J. Perez-calatayud et al., Assessment of the linear reference air kerma rate of Ir-192 wires, PHYS MED BI, 46(8), 2001, pp. 2201-2207
Citations number
10
Categorie Soggetti
Multidisciplinary
Journal title
PHYSICS IN MEDICINE AND BIOLOGY
ISSN journal
00319155 → ACNP
Volume
46
Issue
8
Year of publication
2001
Pages
2201 - 2207
Database
ISI
SICI code
0031-9155(200108)46:8<2201:AOTLRA>2.0.ZU;2-G
Abstract
In this study, a procedure to test the linear reference air kerma rate of I r-192 wires using a well-type chamber is described. The method is based on a special lead insert with a I cm acrylic aperture that provides a differen tial response of the well chamber. The wire is considered divided into I cm parts. Using an external positioning system it is possible to place every part of the wire at the aperture position in the lead insert allowing measu rement of each I cm making up the length of the wire. By means of a set of equations that take into account the contribution of all parts of the wire, in all possible positions, the relative linear reference air kerma rate is obtained. The estimated uncertainties of this procedure are about 2 to 3%. So, a well chamber and the specific inserts allow the measurements of tota l and linear reference air kerma rate for Ir-192 wires.