The new external beam facility of the Oxford scanning proton microprobe

Citation
Gw. Grime et al., The new external beam facility of the Oxford scanning proton microprobe, NUCL INST B, 181, 2001, pp. 66-70
Citations number
9
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS
ISSN journal
0168583X → ACNP
Volume
181
Year of publication
2001
Pages
66 - 70
Database
ISI
SICI code
0168-583X(200107)181:<66:TNEBFO>2.0.ZU;2-I
Abstract
This paper describes the development of a high spatial resolution external beam facility on one of the beamlines of the Oxford scanning proton micropr obe tandem accelerator. Using a ma magnetic quadrupole doublet to focus the beam through the Kapton exit window a beam diameter of < 50 mum full width at half maximum (fwhm) can be achieved on a sample located at 4 mm from th e exit window. The facility is equipped with two Si-Li X-ray detectors for proton-induced X-ray emission (PIXE) analysis of light and trace elements r espectively, a surface barrier detector for Rutherford backscattering spect rometry (RBS) analysis and a HP-Ge detector for gamma -ray detection. The m echanical and beam-optical design of the system is described. (C) 2001 Else vier Science B.V. All rights reserved.