F. Spurny et al., Spectrometry of the LET with track etched detectors - correlation with proportional counter measured spectra, RADIAT MEAS, 34(1-6), 2001, pp. 193-197
A spectrometer of the linear energy transfer (LET) based on the chemically
etched polyallyldiglycolcarbonate (PADC) track etched detector was develope
d. The LET spectra are determined through the measurements of track paramet
ers, it covers LET range between 10 and 700 keV/mum in tissue. A combined e
xperimental and theoretical approach allowed the estimation of the critical
dimensions of the sensitive volume necessary for developing a track to sev
eral nm. It seemed interesting to us to compare the LET spectra obtained by
this method with the microdosimetric spectra available on the basis of a c
lassical experimental microdosimetry method, a tissue equivalent proportion
al counter, for which the critical dimensions simulated are of the order of
a few mum.
Both methods of experimental microdosimetry were compared in the high energ
y radiation reference fields and on the subsonic aircraft board. It was fou
nd out that the microdosimetric distributions are similar; some differences
are, nevertheless, observed. Further studies with the goal to explain them
are outlined. (C) 2001 Elsevier Science Ltd. All rights reserved.