SQUID-based high-resolution thermometer

Authors
Citation
Tcp. Chui, SQUID-based high-resolution thermometer, CRYOGENICS, 41(5-6), 2001, pp. 407-414
Citations number
21
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science","Mechanical Engineering
Journal title
CRYOGENICS
ISSN journal
00112275 → ACNP
Volume
41
Issue
5-6
Year of publication
2001
Pages
407 - 414
Database
ISI
SICI code
0011-2275(200105/06)41:5-6<407:SHT>2.0.ZU;2-Y
Abstract
High-resolution thermometers (HRTs) based on the detection of magnetization changes of a magnetic material using very sensitive superconducting quantu m interference device (SQUID) have routinely achieved temperature resolutio n of 0.1 nK/root Hz at temperatures below 4 K in several laboratories. The resolution of these thermometers has reached the fundamental limit imposed by thermodynamic fluctuations. The principle of operation as well as the th eory of the noise are presented. Some recent discoveries using HRT are disc ussed. (C) 2001 Elsevier Science Ltd. All rights reserved.